ULTRAPENx2™ PTBT5 Specifications:
System Requirements: |
Myron L® Company PTBTX2TM App: Apple iOS 10.0 or later; Android OS 7.0 or later. |
Specifications |
DO Concentration |
DO Saturation |
Temperature |
Ranges |
0-20 ppm |
0-200% |
0 – 71°C / 32 – 160°F |
Accuracy1 |
± 0.2 ppm or ± 2% of the reading, whichever is greater (up to 40ºC) |
0.0% to ≤ 100.0%: ± 20 counts; >100%: ± 2% of the reading. |
± 0.1°C |
Resolution |
0.01 ppm |
0.1% |
0.1°C/F |
Altitude Adjustment: |
-2,000 ft. to 15,000 ft. in increments of 100 ft. |
Salinity Adjustment: |
0 – 50 ppt in increments of 1 ppt |
Time to Reading Stabilization: |
10 – 180 seconds |
Calibration Methods: |
1 or 2 Point; 0 ppm DO Calibration Solution, Air or Water |
Operating/Storage Temperature: |
0 – 55°C / 32 – 131°F |
Power Consumption: |
Active Mode: 30 – 140 mA, Sleep Mode: 2 µA |
Temperature Compensation for Membrane Permeability: |
Automatic |
Physical Dimensions: |
17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D |
Weight: |
54 g or 1.94 oz. |
Case Material: |
Anodized Aircraft Aluminum with Protective Coating |
Battery Type: |
N type, Alkaline 1.5 V |
Water Resistance: |
IP67 and NEMA 6 |
EN61236-1: 2006 – Annex A: 2008; 1999/5/EC Annex III Electrostatic discharge to case of instrument may cause PT5 to spontaneously power ON. In this case, the PT5 will power OFF after several seconds. |
|
|
WARNING! These products can expose you to chemicals including Di(2-ethylhexyl)phthalate (DEHP), which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov. |
1Accuracy after Temperature Controlled, Full Calibration.
ULTRAPENx2™ PTBT5 Measurement Modes:
MODE |
METHOD OF MEASUREMENT |
Hold |
Displays real-time readings until stable or 3 minutes, whichever comes first, then displays final readings. |
LIVE |
Displays real-time readings continuously for up to 5 minutes. |
ULTRAPENx2™ PTBT5 Parameter Modes:
MODE |
PARAMETER MEASURED |
DO Default |
Display alternates between DO Concentration in ppm, DO Saturation in % and Temperature. |
DO Concentration |
Display alternates between DO Concentration in ppm and Temperature. |
DO Saturation |
Display alternates between DO Saturation in % and Temperature. |
Data Sheet