Go Back To:
TESTERS & MONITORS

ULTRAMETER™ III 9PTKB

Ultrameter™ III 9PTKB – Advanced 9-Parameter Water Quality Meter for Accurate Testing | All For Water

ULTRAMETER™ III 9PTKB

ULTRAMETER III 9P™ Hard Carry Case Kit
Display: 4 Digit LCD Outer Material: ABS plastic
Dimensions (LxWxH): 196 x 68 x 64 mm/7.7 x 2.7 x 2.5 in. Inner Material: Foam
Weight: 352 g/12.4 oz. Dimensions: L x W x D: 13.6 x 10 x 5.4 in./346 x 254 x 137 mm
Case/Cell Material: VALOX®* * ® Sabic Innovative Plastics IP BV
Cell Capacities (w/o cell extender): pH/ORP 1,2 mL/0.04 oz. 

 

Cond/TDS/Res: 5 mL/0.2 oz.

 
Power: 9v Alkaline Battery  
Battery Life: 100 Hours/5000 Readings  
Operating/Storage Temperature: 0-55°C/32-132°F  
Protection Ratings: IP67/NEMA 6 (waterproof to 1 meter/3 feet)  
EMI/EMC Ratings (Conformité Européenne): IEC 61326-1: 2020; CISPR 11: 2015 +A1: 2016 +A2: 2019; IEC 61000-4-2: 2008; IEC 61000-4-3: 2010; IEC 61000-4-8: 2009; (EU) 2015/863 (RoHS3): +(EU) 2024/232

Functional Specifications:

Conductivity Ranges: 0–9999 μS/cm; 10–200 mS/cm; in 5 autoranges
  Resolution: 0.01(<100 μS); 0.1(<1000 μS); 1(<10 mS); 0.01(<100 mS); 0.1(<200 mS)
  Accuracy: ±1% of reading
  Auto Temperature Compensation: 0–71°C / 32–160°F
  Adjustable Temperature Compensation to 25°C: 0–9.99%/°C
  Conductivity/TDS Ratios Preprogrammed: KCl, 442*, NaCl
  Adjustable Conductivity/TDS Ratio Factor: 0.20–7.99
TDS Ranges: 0–9999 ppm; 10–200 ppt; in 5 autoranges
  Resolution: 0.01(<100 ppm); 0.1(<1000 ppm); 1(<10 ppt); 0.01(<100 ppt); 0.1(<200 ppt)
  Accuracy: ±1% of reading
  Auto Temperature Compensation: 0–71°C / 32–160°F
  Adjustable Temperature Compensation to 25°C: 0–9.99%/°C
  Conductivity/TDS Ratios Preprogrammed: KCl, 442*, NaCl
  Adjustable Conductivity/TDS Ratio Factor: 0.20–7.99
Resistivity Ranges: 10 KΩ–30 MΩ
  Resolution: 0.01(<100 KΩ); 0.1(<1000 KΩ); 0.1(>1 MΩ)
  Accuracy: ±1% of reading
  Auto Temperature Compensation: 0–71°C / 32–160°F
  Adjustable Temperature Compensation to 25°C: 0–9.99%/°C
pH Ranges: 0–14 pH
  Resolution: 0.01 pH
  Accuracy**: ±0.01 pH
  Auto Temperature Compensation: 0–71°C / 32–160°F
ORP Ranges: ±999 mV
  Resolution: 1 mV
  Accuracy**: ±1 mV
Free Chlorine FCE™ Ranges: 0.00–9.99 ppm when 350≤ORPmV<725 and 0.0≤pH<9.9 and when 725≤ORPmV<825 and 0.0≤pH<8.9
  Resolution: 0.01 ppm
  Accuracy**: <1.00ppm ±0.3ppm and ≥1.00ppm ±0.2ppm
  Auto Temperature Compensation: 0–71°C / 32–160°F
Alkalinity Titration Ranges: 10-800 ppm
  Resolution: 0.1 ppm
Hardness Titration Ranges: 0-1710 ppm / 0-100 grains
  Resolution: 0.1(<1000 ppm); 1(<1710 ppm); 0.1(<100 grains)
LSI Calculator Ranges: -10 to +10
  Resolution: 0.01
Temperature Ranges: 0–71°C / 32–160°F
  Resolution: 0.1°C/F
  Accuracy: ±0.1°C
*442 Natural Water Standard™ Myron L® Company
**EM Susceptibility: When measuring 4.00 pH Buffer in 3.5 molar KCl:
        ± 0.2 pH in the presence or RF fields ≥ 3 V/m.
        ± 0.37 pH in the presence of RF fields ≥ 3 V/m at 300 MHz (± 30 MHz) and 625 MHz (± 50 MHz ).

Data Sheet

Request More Information Now!